6.4.2
Prepare the conductivity probe
Install the conductivity probe
To install a new conductivity probe or to replace a probe follow the instructions below.
Action
Step
Take the probe out of the packaging.
Tip:
Keep the accompanying manual. It will be needed for calibration.
1
Rinse the sensor end of the probe with de-ionized water.
2
Connect the conductivity probe with the probe cable so that no thread is
visible on the probe.
3
NOTICE
Do not cross-thread the probe cable.
In the accompanying manual, locate the cell constant value.
4
Locate the pH/conductivity transmitter display on the I/O Cabinet. Press
MENU (1).
5
1
2
3
Result: A menu is displayed on the transmitter screen.