7.10.2
Remove the disposable bag
Disconnect the disposable bag
Follow the instructions below to disconnect the probes and lines from the disposable
bag after finished mixing procedure.
Action
Step
Make sure the disposable bag is empty.
1
Make sure that the agitator is not running. See
for instructions.
2
Remove the mixer vessel lid.
3
Remove the temperature probe from the disposable bag.
4
Disconnect the pH and conductivity probe cables, and the temperature
probe cable as applicable.
5
Remove and clean the pH and conductivity probes according to facility
procedures. Autoclave and store the probes in a clean place according to
manufacturer's instructions.
6
Clamp all probe connector ends.
7
Disconnect all lines.
8
Xcellerex XDUO 2500 Mixer Operating Instructions 29146667 AH
243
7 Operation
7.10 Finish a batch run
7.10.2 Remove the disposable bag