Action
Step
On the Temperature Detail screen, press the Probe 1/Probe 2 button (1) to
select the applicable sensor.
5
1
2
Make sure that the instrument is in Auto mode (2). If the Manual button is
shown, tap the button to set the pH control to Auto mode.
6
Type an applicable temperature setpoint value into the TC_163 Temperature
SP text box.
7
Result: The control of mixer vessel temperature is enabled.
Note:
Other text boxes on the screen are inaccessible in Auto mode.
226
Xcellerex XDUO 2500 Mixer Operating Instructions 29146667 AH
7 Operation
7.6 Control the temperature