Corrective action
Possible causes
Problem
Verify that the mixer vessel legs
are free of any obstructions or
blocking objects.
Physical obstruction of load cell
motion.
Weight monitoring error:
•
Weight display on local con-
trol software home screen
displays a value outside of
the operational range
or
•
A High High or Low Low
alarm is triggered when the
mixer weight is within the
operational range.
Disengage the jacking bolt. See
Disengage jacking bolts, on
page 147
for instructions.
Note:
If the mixer was filled while the
jacking bolt was engaged, the
weight measurements of the
batch will not be accurate.
Jacking bolt is engaged.
Contact the metrology depart-
ment to do a test (cornering test)
and determine which load cell
has failed.
Load cell has failed (physically
or electrically broken).
Contact your GE representative.
Weight transmitter configuration
incorrect or corrupt.
Xcellerex XDUO 2500 Mixer Operating Instructions 29146667 AH
277
9 Troubleshooting
9.10 Weight monitoring