![EG Xcellerex XDUO 2500 Operating Instructions Manual Download Page 127](http://html1.mh-extra.com/html/eg/xcellerex-xduo-2500/xcellerex-xduo-2500_operating-instructions-manual_2367561127.webp)
Disconnect mixer
The mixer must be turned off and a tag-out device attached onto the I/O Cabinet during
installation. The system power switch must be set in the O OFF position, power cord
must be disconnected or the power supply circuit breaker must be switched off.
Computer
An external computer is not required.
Temperature control unit
For temperature control of the jacketed XDUO 2500 Mixer, a temperature control unit
(TCU) is required. See the manufacturer´s manual for specified environmental conditions
and electrical power requirements for the TCU.
NOTICE
The temperature control unit needs to be disconnected before
moving the equipment.
Connection to FlexFactory
A PROFIBUS cable must be installed between the mixer I/O Cabinet and FlexFactory PLC
Cabinet.
Xcellerex XDUO 2500 Mixer Operating Instructions 29146667 AH
127
5 Installation
5.2 Site requirements