![EG Xcellerex XDUO 2500 Operating Instructions Manual Download Page 140](http://html1.mh-extra.com/html/eg/xcellerex-xduo-2500/xcellerex-xduo-2500_operating-instructions-manual_2367561140.webp)
Action
Step
Turn off the TCU.
1
NOTICE
Do not drain the jacket immediately after use due to
coolant high temperature.
Disconnect the coolant inlet and outlet tubing from the TCU.
2
Prepare a 70 L container for collection of the drained coolant.
3
Connect the drain valve to the collection container as appropriate.
4
Open the drain valve.
5
Open the vent valve and collect the coolant.
6
When the fluid flow stops, lift the coolant outlet tubing (Return) to drain the
content of the tubing into the vessel jacket.
7
Prepare the tubing for draining the vessel jacket bottom:
8
a
Attach a Snap-tite connector to the tubing.
b
Put the end of the drain tubing into the collection container.
c
Connect the Snap-tite connector of the drain tubing to the
coolant inlet tubing (connected to Supply).
See
Fill the vessel jacket, on page 135
for more information about Snap-tite
connector.
Wait until all fluid has drained out from the mixer vessel jacket.
9
140
Xcellerex XDUO 2500 Mixer Operating Instructions 29146667 AH
5 Installation