![EG Xcellerex XDUO 2500 Operating Instructions Manual Download Page 184](http://html1.mh-extra.com/html/eg/xcellerex-xduo-2500/xcellerex-xduo-2500_operating-instructions-manual_2367561184.webp)
Action
Step
Insert the pipette into the probe sheath assembly through the threaded hole
in the probe sheath assembly end plug.
3
Inject the water into the probe sheath assembly.
Note:
The inserted water provides the steam inside the probe sheath assembly
during autoclaving.
4
Remove the pipette.
5
Verify that the O-rings are present on the probe.
6
Insert the probe into the probe sheath assembly through the threaded hole
in the probe sheath assembly end plug.
7
184
Xcellerex XDUO 2500 Mixer Operating Instructions 29146667 AH
6 Preparation
6.4 Prepare and install the probes
6.4.3 Autoclave probe sheath assembly