Action
Step
Hold the probe sheath assembly in its fully compressed state in one hand
and tighten the three ratchet clamps with a tool, until 7 clamp teeth are
engaged.
11
NOTICE
If the ratchet clamps are tightened too much, they can
break. Previous experience has shown that even if a
ratchet clamp breaks, the probe itself would not be
damaged, even if it is a pH probe with a glass body.
Result: The tip of the probe extends above the surface at the inside of the
disposable bag for about 8 mm. The probe is now ready to be connected to
the appropriate cable.
Xcellerex XDUO 2500 Mixer Operating Instructions 29146667 AH
195
6 Preparation
6.4 Prepare and install the probes
6.4.4 Insert probes into the disposable bag