9.2 Instrument Calibration
NOTE:
In this section the word
Calibration
means to set the Instrument to an
internationally recognized standard, independent of input device.
9.2.2 General Calibration Procedure
1.) Use the
u
or
t
keys to show the following functions:
-
Firmware revision
-
Pressure input counts
-
Zero, for the strain gage input (P.ST.ZE)
-
Span, for the strain gage input (P.ST.SP)
-
Pressure (P.STR)
-
Zero, for the linear inputs (P.LN.ZE)
-
Span, for the linear inputs (P.LN.SP)
-
Current (P.020)
-
Voltage, 0-10V (P.010)
-
Secondary input counts
-
Zero, for the strain gage input (S.ST.ZE)
-
Span, for the strain gage input (S.ST.SP)
-
Pressure (S.STR)
-
Zero, for the linear inputs (S.LN.ZE)
-
Span, for the linear inputs (S.LN.SP)
-
Current (S.020)
-
Voltage, 0-10V (S.010)
-
Thermocouple and RTD (S.TC.PT)
-
Reference junction (S.RJ)
-
Line resistance for RTD (S.RL)
-
Line frequency (FREQ)
-
Digital inputs status (DIG.IN)
-
Maximum Power Consumption
-
Minimum Power Consumption
2.) The display values for analog inputs are scaled from 0 to 25,000 counts; it is also
linear for RTD input.
3.) Use the
u
or
t
keys to select a display value from 0 to 10 and to check the linearity
of output circuit at 0%, 10%, 90% and 100%.
4.) If the values do not correspond with the values in the Calibration Parameters
Summary Table below, use the
u
or
t
keys to correct the value displayed.
5.) When all the appropriate values are correct, depress the
FUNC
key.