Indicators (Beacons):
Red LED’s annunciations:
•
A1: Lit when alarm 1 is in alarm state
•
A2: Lit when alarm 2 is in alarm state
•
A3: Lit when alarm 3 is in alarm state
•
REM: Lit when device is controlled by serial link
•
0-25-50-75-100-%: These six LEDs are always on to improve the bar graph
indication.
•
Kg/cm
2
, PSI, Bar, MPa: Engineering unit for the pressure input. These four
beacons are located within the display window, between the numeric digits and the
alarm beacons.
Green LED’s annunciations:
•
PEAK: Lit when lower display shows the peak value
•
TEMP: Lit when lower display shows the temperature input value
2.4 Primary Input
Primary Input:
Selectable between strain gage and linear by software configuration.
Strain Gage Input:
From 340 to 5000
K
, 1-4mV/V. Excitation 10V ±7%. 5 wire
connection.
Input Signal:
-25/125% of full scale (approximately –10 / 50mV).
Shunt Calibration:
With or without shunt resistor (value programmable from 40.0 to
100.0%).
Zero Balance:
±25% of full scale (approximately ± 10mV).
Linear Input:
Selectable between 0-5Vdc, 0-10Vdc, 0-20mA, 4-20mA.
Auxiliary Power supply:
24Vdc / 1.5W ± 2% power supply for two or four wire
transmitter.
Input Impedance:
<10
K
for linear current input
>165K
K
for linear voltage input
Input Protection:
Open circuit detection for strain gage (on signal and excitation wires)
and 4-20mA inputs; it is not available for 0-5Vdc, 0-10Vdc and 0-20mA. Up or down scale
keyboard programmable.
Sampling Time:
50mS typical.
Display Update Time: