DYNISCO UPR800 Series Instruction Manual
Page | 45
Configuration
Parameters are arranged in Groups 5 through 1 and are in order of frequent use. A
general rule of thumb Group 5 parameters are modified once during setup of the indicator
and gradually increase in frequency of use to Group 1 which includes Alarm thresholds
that may be changed on a regular basis.
Parameters that are used only during differential operation are:
Group 1 PI.VAL – Used to display process value on Primary (Main) input
Group 1 SI.VAL – Used to display process value on Secondary input
Group 2 ZER.2.C – Used to zero calibrate secondary analog input
Group 2 SPN.2.C – Used to Span calibrate the secondary analog input
Group 3 SI.FSV – Used to scale secondary input full scale value
Group 3 SI.LSV – Used to scale secondary input lower scale value
Group 4 SI.IFS – Used to set secondary input Failsafe value
Group 5 SI.TYP (Str) – When set to Strain gage, controller sets SI.FNC to diff P
Group 5 SI.FNC – Used to toggle between Temperature or Differential mode
Group 5 – Input Output types
1. Proceed to Group 5 parameters by pressing the
FUNC
button until Group is shown
on lower display and None is shown on upper display. Press the
t
arrow until
Group 5 is displayed.
2. Press
FUNC
button once to display PI.TYP
3. Press
t
or
u
arrow until the desired Primary Input type is selected. Str for Strain
gage, 0-20 for 0-20 mA, 4-20 for 4-20 mA, 0-5 for 0-5VDC, 0-10 for 0-10 VDC.
4. Press
FUNC
button to save and move on to the next parameter.
5. SI.TYP should now be displayed. Press
t
or
u
arrow until the desired Secondary
Input type is selected. Str for Strain gage, 0-20 for 0-20 mA, 4-20 for 4-20 mA, 0-5
for 0-5VDC, 0-10 for 0-10 VDC. Note: Setting to either OFF, tc or RTD will turn off
differential mode.
6. Press
FUNC
button to save and move to the next parameter. SI.FNC will now be
displayed. If Strain gage was selected for SI.TYP, then the SI.FNC will
automatically be set to diFF.P for Differential Pressure otherwise press the
t
arrow to select diFF.P
7. Press
the
FUNC
button to save and move to the next parameter. MO.TYP will be
displayed on the lower display. Press the
t
arrow to select the appropriate Main
Analog Output Type. Analog outputs available are 0-10 for 0-10VDC, -10.10 for -
10V to +10VDC, 0-5 for 0-5VDC, 0-20 for 0-20 mA, 4-20 for 4-20 mA. This
selection should match the remote device analog input type used to remotely
display the Delta Pressure.
8. Press
FUNC
to save and move to the next parameter. SO.TYP will be displayed (if
option is fitted) on the lower display. Press the UP arrow to select the appropriate
Secondary Retransmission Output Type. Retransmission outputs available are 0-
10 for 0-10Vdc, -10.10 for -10V to +10Vdc, 0-5 for 0-5VDC, 0-20 for 0-20mA, 4-20
for 4-20mA or OFF if the Retransmission output is not used. This selection should