Chapter 5 Maintenance and Troubleshooting
Agility System
Operator’s Manual
85
5.2 Cleaning and Decontamination
Warning: The samples, reagents, wash solution and/or the aspirated wash
solution may contain toxic materials or biohazards. The service engineer must
assume that these liquids and any components that come in contact with them
are hazardous.
Ensure that the system has been decontaminated or take suitable precautions
(e.g., wear rubber gloves and suitable eye protection).
If the system is sent back to a Dynex Technologies facility, the user must
certify that the unit has been decontaminated.
5.2.1 Cleaning the System
The system is constructed from materials that are resistant to chemical attack. Spills should be
cleaned up as soon as possible and all external surfaces should be kept clean.
Warning: Make certain that the power cable is disconnected before cleaning
the instrument.
To Clean the System
1.
Clean external surfaces, waste containers, wash containers, consumable tray bays and
tubing lines with a cloth moistened with mild laboratory detergent.
Note:
If needed, dilute the laboratory detergent according to the manufacturer’s
instructions before using.
Note: If there is any doubt about the compatibility of cleaning agents with parts of
the AGILITY system, contact DYNEX Technologies.