Chapter 5 Maintenance and Troubleshooting
90
Agility System
Operator’s Manual
Figure 5-4: Removing the Lamp Assembly
4.
Disconnect the lamp harness to the connector on the Motor Plate.
5.
Remove the old lamp and place a new lamp in position
Note: When replacing the lamp, take care not to get fingerprints or other foreign
material on it. It is recommended that gloves are worn during this procedure. If
fingerprints or other foreign material gets deposited on the lamp, remove it with a
soft tissue saturated with methanol.
6.
After the cap head screws are tightened, place Loctite 425 on them to secure them.