
English
4
Supplement Scio Four modules
Maintenance
Accessories
Connecting to other devices
Device combinations approved by Dräger meet the requirements
of the following standards:
– IEC 60601-1 (3rd edition)
Medical electrical equipment
Part 1: General requirements for basic safety and essential
performance
– IEC 60601-1-2
Medical electrical equipment
Part 1-2: General requirements for safety and essential
performance
Collateral standard: Electromagnetic compatibility –
Requirements and tests
– IEC 60601-1 (2nd edition)
Medical electrical equipment
Part 1: General requirements for safety
– IEC 60601-1-1
Medical electrical equipment
Part 1-1: General requirements for safety
Collateral standard: Safety requirements for medical elec-
trical systems
– IEC 60601-1-2
Medical electrical equipment
Part 1-2: General requirements for safety and essential
performance Collateral standard: Electromagnetic com-
patibility - Requirements and tests
– IEC 60601-1-4
Medical electrical equipment
Part 1-4: General requirements for safety
Collateral standard: Programmable electrical medical sys-
tems
If Dräger devices are connected to other Dräger devices or third-
party devices and the resulting combination is not approved by
Dräger, the correct functioning of the devices may be compro-
mised. The operating organization is responsible for ensuring that
the resulting system meets the requirements of the applicable
standards.
Information on electromagnetic compatibility
General information on electromagnetic compatibility (EMC)
according to international EMC standard IEC 60601-1-2:
Medical electrical equipment is subject to special precautionary
measures concerning electromagnetic compatibility (EMC) and
must be installed and put into operation in accordance with the
EMC information provided in the instructions for use of the patient
monitor.
WARNING
Risk of medical device failure and of patient injury
The medical device must be inspected and serviced regu-
larly by service personnel. Repair and complex mainte-
nance carried out on the medical device must be performed
by specialized service personnel.
If the above is not complied with, medical device failure and
patient injury may occur. Observe chapter "Maintenance".
Dräger strongly recommends that a service contract is
obtained with DrägerService and that all repairs are per-
formed by DrägerService. For maintenance Dräger recom-
mends the use of authentic Dräger repair parts.
WARNING
Risk due to incompatible accessories
Dräger has tested only the compatibility of accessories
listed in the current list of accessories. If other, incompati-
ble accessories are used, there is a risk of patient injury due
to medical device failure.
Dräger recommends that the medical device is only used
together with accessories listed in the current list of acces-
sories.
WARNING
Risk of operating errors and incorrect use
Strictly observe the instructions for use of all accessories.
WARNING
Risk of electric shock and of device failure
Any connected devices or device combinations not com-
plying with the requirements mentioned in these instruc-
tions for use can compromise the correct functioning of the
medical device and lead to an electric shock. Before oper-
ating the medical device, strictly comply with the instruc-
tions for use of all connected devices or device
combinations.
WARNING
Risk of electric shock
Connecting devices to the RS232 or MEDIBUS interfaces
can lead to an increased leakage current. If the protective
ground of one of these devices fails, the leakage current
may rise above the permissible values. Only connect with
the approval of the respective device manufacturer. Have
the leakage current checked by service personnel.
If the permissible value is exceeded, disconnect the
devices from the RS232 or MEDIBUS interface.
WARNING
Risk of operating errors and incorrect use
Strictly observe assembly instructions and instructions for
use for each connected device.
WARNING
Risk of device failure
Electromagnetic fields e.g., those generated by radio fre-
quency communication equipment such as mobile phones,
high-frequency electrical surgery equipment, defibrillators
or shortwave therapy devices can disrupt the function of
the medical device.
Only operate radio frequency devices at a sufficient safety
clearance.
Summary of Contents for Scio Four
Page 19: ...Supplement Scio Four modules 19 This page intentionally left blank...
Page 23: ...Scio Four 23 MC IEC 60601 1 2 MC MC MC MEDIBUS 0 3 1 0 MEDIBUS 38 Dr ger...
Page 27: ...Scio Four 27 1 RS232C X2 A 2 Infinity 1 HME Luer Lock 2 1 2 002 A...
Page 28: ...28 Scio Four 1 2 1 2 3 1 1 2 HME...
Page 30: ...30 Scio Four 1 2 3 4 1 2 3 4 Dr ger Dr gerService 1 1 12 10...
Page 39: ...Scio Four 39...
Page 129: ...Dodatak Scio Four moduli 129 Ova je stranica namjerno ostavljena prazna...
Page 151: ...Scio Four 151 MEDIBUS Scio Four CO2 N2O O2 0 3 m 1 0 MEDIBUS 166 Dr ger PVC...
Page 155: ...Scio Four 155 1 Y Luer Lock 2 1 2 AGSS 1 2 O 1 2...
Page 156: ...156 Scio Four 1 2 3 1 Dismozon plus Bode Chemie...
Page 163: ...Scio Four 163 CISPR 11 1 A IEC 61000 3 2 IEC 61000 3 3...
Page 167: ...Scio Four 167...
Page 171: ...Scio Four 171 MEDIBUS Scio Four CO2 N2O O2 0 3 m 1 0 ft MEDIBUS 186 Dr ger PVC...
Page 172: ...172 Scio Four 25 MRI NMR NMI 25 cm 10 in Dr ger 001 B 002 B MEDIBUS H RS232C X2 B B H...
Page 175: ...Scio Four 175 1 Luer Lock 2 1 2 1 2...
Page 176: ...176 Scio Four 1 2 1 2 3 1...
Page 187: ...Scio Four 187...
Page 205: ...Ek Scio Four mod lleri 205 Bu sayfa kasten bo b rak lm t r...
Page 212: ...212 Scio Four 1 Y HME Luer 2 1 2 1 On Off ON 2 O 1 2...
Page 229: ...Scio Four 229 1 ON OFF ON 2 1 2 3 OFF 1 ON OFF OFF 1 2 HME...
Page 239: ...Scio Four 239...