NM2B
REPLACEMENT PROCEDURES (continued)
4.19.22 Attach the large hose from the muffler interface wall and secure the
connection with a new tie strap.
4.19.23 Attach the other large hose to the exhaust port on the valve case
assembly and secure the connection with a new tie strap.
4.19.24 Remove the prop from the monitor box and carefully lower it into
position. Replace the screws holding the monitor box to the ventilator
box.
4.19.25 Place the bellows box front panel into position, ensure that the slot
in the knob assembly is correctly aligned with the drive pin on the
bellows adjustment shaft, and install the four screws holding the
front panel to the machine.
4.19.26 Reattach the canister to the bellows box.
4.19.27 Reattach the bellows assembly and tighten the wingnuts holding it
in place. Connect any hoses that were previously removed from the
bellows assembly.
4.19.28 Reconnect the pipeline hoses and AC power cord, and reset all circuit
breakers.
4.19.29 Perform the PMS Procedure given in Section 6.
Rev. A
4-63