PMC PROCEDURE (continued)
NM2B
( )
6.15.3
Air Flowmeter - if applicable
*
6.15.3.1
Set the gas selector switch to ALL GASES.
**
6.15.3.2
Open the air cylinder valve.
**
6.15.3.3
Is it possible to adjust the flow of the air over the full range of the
flowmeter? ___ (Y)
**
6.15.3.4
Close the air cylinder valve and bleed the pressure.
6.15.3.5
Close the air flow control valve.
6.15.3.6
Connect the Air pipeline supply and verify the proper operation of
the air flowmeter. Close the air flow control valve.
6.15.3.7
Is the correct flow control knob and label attached to the air flow
control valve? ___ (Y)
*
6.15.3.8
What is the minimum flow of oxygen? ___ ml (0)
( )
6.15.4
Oxygen-Helium Flowmeter - if applicable
*
6.15.4.1
Set the gas selector switch to ALL GASES.
6.15.4.2
Open the oxygen-helium gas cylinder valve.
6.15.4.3
Is it possible to adjust the flow of the oxygen-helium over the full
range of the flowmeter? ___ (Y)
6.15.4.4
Close the oxygen-helium flow control valve.
6.15.4.5
Is the correct flow control knob and label attached to the oxygen-
helium flow control valve? ___ (Y)
* These items do not apply to machines with later design (P/N 4113329-001) ORC.
**these items do not apply to 4-gas machines.
Rev. Y
6-36
Summary of Contents for Narkomed 2B
Page 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 18: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 32: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 240: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 244: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 306: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...