PMC PROCEDURE (continued)
NM2B
6.17.5
Nitrogen Concentration - if applicable
6.17.5.1
Depress the O
2
FLUSH button for 5 seconds.
6.17.5.2
Does the O
2
MED read 97-100% within 3 minutes?___ (Y)
6.17.5.3
Set the gas selector switch to ALL GASES.
6.17.5.4
Set O
2
flow to 4 l/min; nitrogen flow to 2 l/min.
( )
6.17.5.5
What is the oxygen concentration after the value stabilizes? ___ %
(64-70)
6.17.5.6
Close the nitrogen cylinder valve and bleed the pressure from the
circuit.
6.17.5.7
Close the nitrogen flow control valve.
6.17.5.8
Set the gas selector switch to O
2
+N
2
O.
6.17.5.9
Close the oxygen flow control valve.
6.17.6
Helium Concentration - if applicable
6.17.6.1
Depress the O
2
FLUSH button for 15 seconds.
6.17.6.2
Does the O
2
MED read 97-100% within 3 minutes? ___ (Y)
6.17.6.3
Set the gas selector switch to ALL GASES.
6.17.6.4
Set O
2
flow to 4 l/min; helium flow to 2 l/min.
( )
6.17.6.5
What is the oxygen concentration after the value stabilizes? ___%
(64-70)
6.17.6.6
Close the helium cylinder valve and bleed the pressure from the
circuit.
6.17.6.7
Close the helium flow control valve.
6.17.6.8
Set the gas selector switch to O
2
+N
2
O.
6.17.6.9
Close the oxygen flow control valve.
Rev. Y
6-42
Summary of Contents for Narkomed 2B
Page 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 18: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 32: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 240: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 244: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 306: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...