PMC PROCEDURE (continued)
NM2B
6.26 TRACE & TREND
( )
6.26.1
Trace:
Press the TRACE key and verify that the lower half of the
alarm CRT displays the Breathing Pressure waveform.
( )
6.26.2
Trend:
Press the TREND key and verify that the lower half of the
alarm CRT can selectively display percent oxygen, breaths
per minute, and minute volume.
NOTE:
Information must be available in order to be displayed.
6.26.2.1
Turn the ventilator OFF.
6.27 PRESSURE LIMIT CONTROLLER (AV-2, or optional PLC, if applicable)
6.27.1
Turn the ventilator ON.
6.27.2
Set the Inspiratory Flow to the bottom of the low range.
6.27.3
Set the oxygen flow rate to 4 l/min.
6.27.4
Set the Pressure Limit Control to the MAX position.
6.27.5
Occlude the Y-piece with your thumb.
( )
6.27.6
Slowly increase the Inspiratory Flow setting until a peak pressure of 80
cm H
2
O is achieved.
6.27.7
Set the Pressure Limit Control to 30.
( )
6.27.8
What is the peak pressure? ___ cm H
2
O (27-33)
6.27.9
Set the Pressure Limit Control to the Min position.
( )
6.27.10
What is the peak pressure? ___ cm H
2
O (<15)
6.27.11
Remove your thumb from the Y-piece.
6.27.12
Set the Inspiratory Flow to the maximum of the LOW zone.
6.27.13
Close the oxygen flow control valve.
6.27.14
Turn the ventilator OFF.
Rev. Y
6-58
Summary of Contents for Narkomed 2B
Page 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 18: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 32: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 240: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 244: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 306: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...