NM2B
PMC PROCEDURE (continued)
( )
6.13.3
Air High Pressure Leak - if applicable
*
6.13.3.1
Set the gas selector switch to ALL GASES.
6.13.3.2
Open the air cylinder valve.
6.13.3.3
Let the pressure stabilize.
6.13.3.4
Close the air cylinder valve and remove the cylinder.
6.13.3.5
Observe the air cylinder pressure gauge.
6.13.3.6
After two (2) minutes, what is the pressure loss? ___ PSI (<50)
6.13.3.7
Attach the air cylinder.
( )
6.13.4
Oxygen-Helium High Pressure Leak - if applicable
*
6.13.4.1
Set the gas selector switch to ALL GASES.
6.13.4.2
Open the oxygen-helium cylinder valve.
6.13.4.3
Let the pressure stabilize.
6.13.4.4
Close the oxygen-helium cylinder valve and remove the cylinder.
6.13.4.5
Observe the oxygen-helium cylinder pressure gauge.
6.13.4.6
After two (2) minutes, what is the pressure loss? ___ PSI (<50)
6.13.4.7
Attach the oxygen-helium cylinder.
( )
6.13.5
Carbon Dioxide High Pressure Leak - if applicable
*
6.13.5.1
Set the gas selector switch to ALL GASES.
6.13.5.2
Open the carbon dioxide cylinder valve.
6.13.5.3
Let the pressure stabilize.
6.13.5.4
Close the carbon dioxide cylinder valve and remove the cylinder.
* These items do not apply to machines with later design (P/N 4113329-001) ORC.
Rev. Y
6-29
Summary of Contents for Narkomed 2B
Page 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 18: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 32: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 240: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 244: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 306: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...