REPLACEMENT PROCEDURES (continued)
NM2B
4.4.14.1
Open the cylinder valve and check for a pressure
indication on the corresponding gauge at the gas
instrumentation panel.
NOTE:
The cylinder used for this test must
contain the following minimum pressure:
O
2
: 1000 PSI
N
2
O
:
700 PSI
HE
: 1000 PSI
CO
2
:
800 PSI
AIR
: 1000 PSI
N
2
: 1000 PSI
O
2
-He : 1000 PSI
4.4.14.2
Close the cylinder valve and remove the cylinder from the
yoke.
4.4.14.3
For any gas, the pressure should not drop more than 50
PSI in two minutes.
4.4.15
Reinstall the cylinder in the yoke.
4.4.16
Place the plexiglass cover over the gauges and flow tubes, and
reinstall the cover screws. Do not over-tighten these screws as the
plexiglass may crack.
4.4.17
Place the knob guard over the flow control valves and install its two
retaining screws. (Reinstall the bar-type knob guard on earlier
machines.)
4.4.18
Reinstall the O
2
flow control knob and tighten its setscrews. If the
knob is installed properly, the O
2
label will be straight when the knob
is against its clockwise stop.
4.4.19
Replace the front plate at the top of the plexiglass cover and secure
it with the hardware that was previously removed.
4.4.20
Replace the rear cover and its retaining screws.
4.4.21
Replace the table top and its retaining screws.
4.4.22
Connect the pipeline hoses.
4.4.23
Perform the PMS Procedure given in Section 6.
Rev. A
4-14
Summary of Contents for Narkomed 2B
Page 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 18: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 32: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 240: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 244: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 306: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...