NM2B
PMC PROCEDURE (continued)
6.16.17
What is the high oxygen alarm default? ___ % (100)
6.16.18
Verify that the high alarm limit has a range from 100 to 22%.
6.16.19
Set the high alarm to 95.
6.16.20
the alarm CRT shall display "% OXYGEN HI" under the ADVISORY
column.
6.16.21
Return the high alarm to 100.
6.16.22
The "% OXYGEN HI" message will now disappear.
( )
6.16.23
Within three minutes, what is the oxygen concentration? ___% (97-100)
Rev. Y
6-39