ADJUSTMENT AND CALIBRATION PROCEDURES (continued)
NM2B
CONNECTION
ON SENSOR
INTERFACE
PANEL
ABSORBER
ASSEMBLY
TOP VIEW
TEST
FIXTURE
BREATHING
PRESSURE
SV20322
Figure 5-7: BAROMED CALIBRATION SETUP AND SCREEN
CALIBRATION MENU
O2MED
OFFSET O2 CELL A
0138
0127
OFFSET O2 CELL B
0138
0129
BAROMED
OXYGEN OFFSET READINGS STORED
PRESSURE SPAN (60 cm H2O)
0243
0481
EXIT
TRACE
TREND
Rev. A
5-22