Service
INDICATOR DG8000 RevB0
94
RESULT
CAUSE
ACTION
Zero stable and correct
functioning
The sensor cable and the
junction box are
NOT
damaged
Proceed with the other tests
Functioning not correct only in
some junction box connectors.
The junction is damaged or wet
Try to dry the junction box and
repeat the test; in case you do
not have success, replace the
junction box.
Zero NOT stable or NOT correct
functioning in all the box’s
connectors
The sensor cable is damaged
Replace the sensors’ cable
Check the working of the SENSORS
a) Open the JUNCTION BOX
b) Leave one sensor and the cable connected the scale
c) Do the TARE (use the microcomputer’s for instructions)
d) The scale must steady, viewing “0” Kg.
e) Verify the right working, trying to load weight on the connected sensor (the displayed weight is not
important, but it must be steady).
REPEAT THE TEST CONNECTING ONE AT EACH THE SENSORS.
RESULT
CAUSE
ACTION
Zero and weight stable.
The sensor is
NOT
damaged
Go on with the other sensors
Zero and weight not stable.
The sensor
is
damaged
Contact the assistance service