Between patients
1.
Remove any ultrasonic inserts and the Steri-Mate
®
handpiece. Clean and sterilize the handpiece and all inserts used
during the procedure (refer to the Infection Control information
booklet and the Cavitron Ultrasonic Inserts Directions for Use
for cleaning and sterilizing instructions).
2. Clean and disinfect the handpiece cable assembly by applying
an approved non-immersion type disinfectant solution*
carefully following the instructions provided by the solution
manufacturer. To clean the cable, generously spray disinfectant
solution on a clean towel and wipe the cable and connector.
Discard used towel. To disinfect the system, generously spray
disinfectant on a clean towel and wipe the cable and connector.
Allow the disinfectant solution to air dry.
3. Clean and disinfect all surfaces of the dental treatment control
head according to the manufacturer’s instructions.
4. Connect a sterilized Steri-Mate
®
handpiece onto its
mating cable connector.
5. Hold the handpiece (without an insert installed) over a sink
or drain. Activate the foot control and flush water through
handpiece for at least thirty seconds.
6. Place a sterilized insert into the handpiece.
Shut down at the end of the day
1. Flush the DENTSPLY Cavitron
®
ultrasonic system
according the dental treatment unit manufacturer’s
instructions.
2. Remove any ultrasonic inserts and the Steri-Mate
®
handpiece. Clean and sterilize the handpiece and all inserts
used during the procedure (refer to the Infection Control
information booklet and the Cavitron Ultrasonic Inserts
Directions for Use for cleaning and sterilizing instructions).
3. Clean and disinfect the handpiece cable assembly by
applying an approved non-immersion type disinfectant
solution* carefully following the instructions provided by
the solution manufacturer. To clean the cable, generously
spray disinfectant solution on a clean towel and wipe the
cable and connector. Discard used towel. To disinfect the
system, generously spray disinfectant on a clean towel and
wipe the cable and connector. Allow the disinfectant solution
to air dry.
4. Clean and disinfect all surfaces of the dental treatment
control head according to the manufacturer’s instructions.
5. Place the cable connector into the handpiece holder for
storage.
6. Turn OFF the power and water to the dental treatment unit.
*NOTE: Water-based disinfection solutions are preferred. Some alcohol-
based disinfectant solutions may discolor the handpiece cable
and connector.
Summary of Contents for Cavitron SPS G90
Page 2: ......
Page 17: ...ENGLISH 15...
Page 18: ...ENGLISH 16...
Page 19: ...ENGLISH 17...
Page 20: ...ENGLISH 18 Optional...
Page 36: ...FRAN AIS 34...
Page 37: ...FRAN AIS 35...
Page 38: ...FRAN AIS 36...
Page 39: ...FRAN AIS 37 En option...
Page 55: ...ESPA OL 53...
Page 56: ...ESPA OL 54...
Page 57: ...ESPA OL 55...
Page 58: ...ESPA OL 56 Opcional...
Page 74: ...DEUTSCH 72...
Page 75: ...DEUTSCH 73...
Page 76: ...DEUTSCH 74...
Page 77: ...DEUTSCH 75 Wahlweise...
Page 93: ...ITALIANO 91...
Page 94: ...ITALIANO 92...
Page 95: ...ITALIANO 93...
Page 96: ...ITALIANO 94 Facoltativa...
Page 101: ...99 4 5 9 Sterile Lavage Kit P N 81340 5 0 12...
Page 105: ...103 6 2 25 172 60 414 60 Cavitron 25 C 77 F Cavitron 7 SPS SPS 7 1 DENTSPLY Cavitron SPS...
Page 106: ...104 7 2 Steri Mate Cavitron 30K Steri Mate...
Page 107: ...105 DENTSPLY 7346 30K FSI SLI 10S 7 3 DENTSPLY Cavitron 30K DENTSPLY Cavitron...
Page 108: ...106 8 Steri Mate DENTSPLY Cavitron 9 9 1 High Vacuum Evacuator HVE 9 2 Cavitron 7 4...
Page 110: ...108 2 3 4 Steri Mate 5 30 6 1 Cavitron 2 Steri Mate Cavitron 3 4 5 6...
Page 111: ...109 11...
Page 112: ...110...
Page 113: ...111...
Page 114: ...112...
Page 115: ...113...
Page 117: ......