ENGLISH • 12
• If sensitivity persists, decrease power setting and/or move
from the sensitive tooth to another and then return.
Section 10: System Care
10.1
Daily Protocol
Start-up procedures at the beginning of the day
1. Switch ON the dental treatment unit (refer to instruction manual
of treatment unit).
2. Remove the DENTSPLY Cavitron
®
cable connector assembly
from handpiece holder and set power adjustment to 30% of
maximum.
3. Connect Steri-Mate
®
handpiece to the cable connector assembly
(refer to the Infection Control information booklet for sterilizing
instructions).
4. Adjust the lavage control to maximum.
5. Hold the handpiece (without an insert installed) over a sink
or drain. Activate the foot control and flush water through
handpiece for at least two minutes.
6.
Place a sterilized insert into the handpiece using a gentle push-
twist motion. Refer to the Cavitron Ultrasonic Inserts Directions
for Use for cleaning and sterilizing instructions.
7. Activate the ultrasonics and adjust the power and lavage control
to your preferred operating positions.
accidental prolonged contact with the activated tip
whenever the insert is placed in the mouth.
• Always adjust the lavage so adequate fluid will be
available to cool the tip-tooth interface.
• In general, it is suggested that a “feather light touch” be
used both supra- and subgingivally. The motion of the
activated tip and acoustic effects of the irrigating fluid,
in most cases, is adequate to remove even the most
tenacious calculus.
• Set the System’s power adjustment knob to the lowest
power setting for the application and the selected insert.
• Only adjust the System’s power adjustment knob with
the insert outside the patients mouth.
9.3 Patient Comfort Considerations
Reasons for sensitivity
• Incorrect tip placement. Point should be directed away
from root surfaces.
• Not keeping the tip in motion on the tooth. Do not allow
the insert to remain in a static position on any one area
of the tooth. Change the insert’s path of motion.
• Applying pressure. Use extremely light grasp and
pressure, especially on exposed cementum.
Summary of Contents for Cavitron SPS G90
Page 2: ......
Page 17: ...ENGLISH 15...
Page 18: ...ENGLISH 16...
Page 19: ...ENGLISH 17...
Page 20: ...ENGLISH 18 Optional...
Page 36: ...FRAN AIS 34...
Page 37: ...FRAN AIS 35...
Page 38: ...FRAN AIS 36...
Page 39: ...FRAN AIS 37 En option...
Page 55: ...ESPA OL 53...
Page 56: ...ESPA OL 54...
Page 57: ...ESPA OL 55...
Page 58: ...ESPA OL 56 Opcional...
Page 74: ...DEUTSCH 72...
Page 75: ...DEUTSCH 73...
Page 76: ...DEUTSCH 74...
Page 77: ...DEUTSCH 75 Wahlweise...
Page 93: ...ITALIANO 91...
Page 94: ...ITALIANO 92...
Page 95: ...ITALIANO 93...
Page 96: ...ITALIANO 94 Facoltativa...
Page 101: ...99 4 5 9 Sterile Lavage Kit P N 81340 5 0 12...
Page 105: ...103 6 2 25 172 60 414 60 Cavitron 25 C 77 F Cavitron 7 SPS SPS 7 1 DENTSPLY Cavitron SPS...
Page 106: ...104 7 2 Steri Mate Cavitron 30K Steri Mate...
Page 107: ...105 DENTSPLY 7346 30K FSI SLI 10S 7 3 DENTSPLY Cavitron 30K DENTSPLY Cavitron...
Page 108: ...106 8 Steri Mate DENTSPLY Cavitron 9 9 1 High Vacuum Evacuator HVE 9 2 Cavitron 7 4...
Page 110: ...108 2 3 4 Steri Mate 5 30 6 1 Cavitron 2 Steri Mate Cavitron 3 4 5 6...
Page 111: ...109 11...
Page 112: ...110...
Page 113: ...111...
Page 114: ...112...
Page 115: ...113...
Page 117: ......