Operating Instructions SiroLaser Blue
4 Installation
4.4 Initial start-up – procedure for proper assembly
66 56 842 D3648
D3648.201.03.02.02 10.2018
31
Bending of the EasyTip
WARNING
For surgical applications, please bend the fiber tip not more than 60°.
1.
Please sterilize the bending tool prior to each use in order to keep
the sterile conditions of the EasyTip, see chapter "Cleaning,
disinfection and sterilization [→ 63]"
2.
After sterilizing the bending tool, you can insert the EasyTip into the
bending tool and bend the EasyTip to the angle that you need for
best handling.
Adjusting the position of the endo stopper
The single-use fiber tip for endodontic applications (EasyTip Endo) is
provided with a pre-mounted endo stopper, which serves as a stop
collar for the length of the root canal. The endo stopper can be moved
on the optical fiber.
1.
To guarantee the sterile conditions of the fiber tip, please use the
sterilized bending tool to adjust the position of the endo stopper.
2.
Place the EasyTip in the bending tool so that the endo stopper
positions itself in the notch for the stopper in the bending tool.
3.
Adjust the position of the endo stopper by moving the handpiece up
and down so that the end of the fiber tip has the needed distance to
the stopper (indicated by the mm scale on the bending tool).
Adjusting the fiber length with the fiber cutter
Generally, all types of sterile delivered single-use fiber tips have the
proper fiber length in order to start working immediately without the
need of adjusting the length prior to application. However, in some
cases it will be necessary to adjust the length of the fiber.
Please sterilize the fiber cutter prior to each use to keep the sterile
conditions of the EasyTip, see chapter "Cleaning, disinfection and
sterilization [→ 63]".