7 Cleaning, disinfection and sterilization
7.4 Cleaning the control unit
66 56 842 D3648
D3648.201.03.02.02 10.2018
65
NOTE
Sterilize the therapy light guides in a packing material suitable for
sterilization and storage to prevent scratching or chipping the light
guides inthe autoclave. Do not exceed a temperature of 140 °C
(284 °F) during the drying cycle. Do not abort the drying cycle before
it has ended. Do not try to accelerate the cool-down process by
placing the MultiTips in cold water. This could cause the therapy light
guide to crack.
CAUTION
Store all components so that they are protected against
contamination. Sterilize again once the storage period has elapsed.
This is not valid for sterile delivered single-use EasyTips.
7.4
Cleaning the control unit
Use a dry, soft cloth to remove dust from the SiroLaser Blue. More
stubborn spots can be removed with a damp cloth.
NOTE
Please proceed carefully not to scratch and damage the foil on the
touch screen.
You can wipe-disinfect the SiroLaser Blue using any of the products that
are commonly used to disinfect medical electrical equipment, e.g.
MinuteWipes, Caviwipes.
CAUTION
Spray disinfection may allow liquids to penetrate into the SiroLaser
Blue!
The SiroLaser Blue may be disinfected only by wiping it. Do not spray
disinfect the SiroLaser Blue laser unit.
Observe the instructions provided by the manufacturers of these
disinfectants.
MinuteWipes Fa. Alpro.
In USA: Caviwipes
™
.