Wireless foot control
Frequency:
2.4 GHz - 2.4835 GHz (ISM band)
Transmitting power:
< 2 mW (short-range device)
Modulation type:
Multi-dimensional Multi Access (MDMA)
Battery:
Type AAA, 1.5 V
Transport and storage
The SiroLaser Blue comes in a cardboard box that ensures proper and
easy transport.
CAUTION
Do not leave the SiroLaser Blue in a vehicle parked in the sun. The
inside temperature of the car could thus heat up to a point where
individual components may be damaged.
To ensure appropriate protective storage, the device must always be
kept in the box supplied by Dentsply Sirona.
Thus stored, the SiroLaser Blue can withstand the following ambient
conditions:
● Temperatures from -40 °C to +70 °C
● Relative humidity from 10 % to 95 %
● Atmospheric pressure from 50 kPa to 106 kPa
NOTE
The rechargeable battery must be fully charged regularly. After six
months of no charging (storage) the rechargeable battery might lose
its loading capacity and might not be rechargeable anymore.
Operating conditions
The SiroLaser Blue may be operated in the following environmental
conditions:
● Temperatures from 50 °F (10 °C) to 91 °F (33 °C)
● Relative humidity from 10 % to 95 %
● Atmospheric pressure from 80 kPa to 106 kPa
CAUTION
Following transport and storage, allow the SiroLaser Blue adapt to
room temperature for one hour prior to operation to reduce the risk of
malfunctions caused by condensation.