2 Safety information
2.10 Contamination
Dentsply Sirona
Operating Instructions SiroLaser Blue
12
66 56 842 D3648
D3648.201.03.02.02 10.2018
2.10
Contamination
Accessories
Danger of (cross) contamination. Pay attention not to hurt or stick
yourself or any other person with the laser fiber tip. This applies also if
the handpiece is placed in the holder.
Prior to each use, the handpiece sleeve and the optical light guide
(MultiTip) must be sterilized. The single-use fiber tips are delivered
sterile and must be used only once and disposed after use.
Tissue particles
During cutting and coagulation of tissue, tissue particles could become
aerosolized. Always wear a face mask, because a risk of infection
exists.
A extractor or a filter should be used. The operating personnel should
be aware that biologically active material could get into the environment.
It may contain particles of viable tissue.
WARNING
Single-use fiber tips (EasyTip) must not be sterilized again after
usage. They are disposable products and are not for re-use.
➢ Never reuse the single-use fiber tips as they can also damage the
handpiece optics if used more than one time.
2.11
Installation
Location
The SiroLaser Blue is to be protected against the intrusion of liquids.
The SiroLaser Blue must not be used in areas in which the presence of
liquids is probable.
Verify that the line voltage corresponds to the voltage indicated on the
rating plate of the power supply or in the technical specifications.
Make sure that the electrical system is equipped with the required
devices for protection against direct and indirect contact
(thermomagnetic switches, residual current circuit breakers) and has
been set up by a qualified electrician in compliance with the applicable
standards.
Avoid interference between the laser emission and any optical sensors
of devices operated in the vicinity of the SiroLaser Blue.
National directives regarding electrical installations must be observed.
Set up
Set up the SiroLaser Blue unit properly and completely before putting it
into operation, see chapter "Installation [→ 25]".