5 Operation
5.5 Main home screen
Dentsply Sirona
Operating Instructions SiroLaser Blue
58
66 56 842 D3648
D3648.201.03.02.02 10.2018
The unit asks that the fin-
ger switch is pressed for
three seconds. This
screen is displayed during
laser calibration for the
calibration process (with-
out power meter)
Confirmation via ‘OK’
The unit asks to direct the
laser beam to the power
meter and to press the fin-
ger switch for three sec-
onds. This screen is dis-
played during laser cali-
bration for the calibration
process (with power me-
ter).
The operator presses
‘Yes’ or ‘No’ if the mea-
sured value in inside the
stated value or not.
5.5.4.3
Information messages
This screen occurs when
the battery level is low and
the external power supply
is not connected.
The device displays the in-
formation that the battery
charge is sufficient only for
180sec of treatment (with
max power) and recom-
mend to connect the
power supply.
The calibration check was
successfully performed.
Confirmation via 'OK'.