![Dentsply Sirona SiroLaser Blue Operating Instructions Manual Download Page 56](http://html1.mh-extra.com/html/dentsply-sirona/sirolaser-blue/sirolaser-blue_operating-instructions-manual_2483705056.webp)
5 Operation
5.5 Main home screen
Dentsply Sirona
Operating Instructions SiroLaser Blue
56
66 56 842 D3648
D3648.201.03.02.02 10.2018
Displayed when there is a
finger switch error.
Please contact Dentsply
Sirona, your local dental
dealer or your authorized
service center.
Confirmation via ‘OK’.
Displayed when an error
of the fan occurs (for ex-
ample: the fan is blocked).
To avoid damage, please
switch off the SiroLaser
Blue and allow it cool
down. See chapter "Trou-
bleshooting of simple de-
fects [→ 68]".
Confirmation via ‘OK’.
Displayed when the inter-
lock contact is open.
See chapter "Trou-
bleshooting of simple de-
fects [→ 68]".
Displayed when the diode
current differs by more
than the 20% tolerance
compared to the calibrated
current.
Please contact Dentsply
Sirona, your local dental
dealer or your authorized
service center.
Confirmation via ‘OK’.
Displayed in case when
there is no fiber con-
nected. See chapter
"Troubleshooting of simple
defects [→ 68]".
Confirmation via ‘OK’.
The battery level is low and
needs to be connected to
the power supply.
Confirmation via ‘OK’.
Device error has occurred.
Please contact Dentsply
Sirona, your local dental
dealer or your authorized
service center.
Confirmation via ‘OK’.
An error occurred during
calibration check (with or
without power meter).
which could not be han-
dled.
Please contact Dentsply
Sirona, your local dental
dealer or your authorized
service center.
Confirmation via 'OK'.