Dentsply Sirona
Operating Instructions SiroLaser Blue
5 Operation
5.5 Main home screen
66 56 842 D3648
D3648.201.03.02.02 10.2018
45
Putting the laser in ready mode
1.
Select the desired preset.
Ä
The pre-set parameters will be displayed.
WARNING
Check the set parameters before activating the treatment.
2.
Now you can activate the laser: Press the 'Activate Laser' button.
Ä
You will be notified to wear the correct protective goggles
before the aiming beam is activated.
3.
Acknowledging the notification.
Ä
The green LEDs start flashing. After a delay of 2 seconds, the
aiming beam is switched on.
Ä
The laser is now ready for operation.
NOTE
If the finger switch or foot switch is actuated during the 2 second
period before the laser is in ready mode, an error message will be
displayed.
WARNING
All persons present in the room (operator, assistants and patient)
must wear the proper laser protective goggles as soon as the laser
unit's notification to wear laser protective goggles appears and
whenever the green LEDs are lit.
Any actuation of the finger or wireless foot control activates the laser
unit.
Incorrect settings may result in severe damage of the patient's soft or
hard tissue or may result in lack of treatment efficacy. Users of this
device must have proper knowledge and training in laser therapy.
The treatment room must be protected by suitable measures (in
compliance with IEC 60825-1), e.g. by closing the doors.
NOTE
Before starting a laser treatment using battery power, please
reconfirm the battery status.
When you actuate the finger or wireless foot control, the laser output is
activated. At the same time, two yellow LEDs at the upper right and left
end of the SiroLaser Blue control unit light up, as does the ’laser active’
bar on the touch screen and the audible alarm sounds. When you
release the finger or wireless foot control to interrupt treatment, the
laser output is deactivated, but the laser remains ready for operation.
The following is a typical example of a treatment submenu.