4 Installation
4.4 Initial start-up – procedure for proper assembly
Dentsply Sirona
Operating Instructions SiroLaser Blue
28
66 56 842 D3648
D3648.201.03.02.02 10.2018
4.4.2
Handpiece and assembly of single-use fiber tips and
therapy rods
4.4.2.1
Handpiece
A
B
D
A
C
A
Handpiece body with tube
B
Snap tab
C
Stainless steel handpiece sleeve
D
Keypad for finger switch
4.4.2.2
Assembly of sterile single-use optical fiber tips
4.4.2.2.1
Area of application
The SiroLaser Blue is provided with three types of sterile single-use
optical fiber tips of different diameter so that it can be used for a variety
of different dental procedures and indications:
● Single-use optical fiber tip, EasyTip 320 (sterile)
● Single-use optical fiber tip, EasyTip 200 (sterile)
● Single-use optical fiber tip, EasyTip Endo (sterile)
For the specification of each type of EasyTip see chapter „Technical
Data [→ 21]“.
EasyTips are delivered sterile in a special packaging tube, which also
assists the mounting of the fiber tips.The optical single-use fiber tips can
be used only with the SiroLaser Blue in the spectral range of 445 nm ±
5 nm and 970 nm -10/+15 nm.
WARNING
If optical fiber tips from other manufacturers are used, physical
properties such as load carrying capacity and transmission behavior
may vary. Dentsply Sirona therefore assumes no liability in such
cases.Therefore, use only Dentsply Sirona single-use optical fiber
tips.
WARNING
The fiber tips from the demo set of single-use fiber tips in the scope of
supply serve only as test fibers for first familarisation with the unit.
They are not sterile, as shown on their label. Do not use them for
clinical treatments of patients.