8 Maintenance and service
8.4 Troubleshooting of simple defects
Dentsply Sirona
Operating Instructions SiroLaser Blue
68
66 56 842 D3648
D3648.201.03.02.02 10.2018
8.4
Troubleshooting of simple defects
In case of malfunctioning, proceed as follows:
General
Initial general directions in case of malfunctioning:
● Check if power supply is connected properly and/or rechargeable
battery is loaded.
● Check if EasyTip or MultiTip is installed correctly.
● Be sure that all operational steps have been carried out correctly.
● Check the functioning of the finger switch and/or wireless foot
control by pressing several times.
If the touch screen of the SiroLaser Blue remains dark after switching it
on:
● Check the connection of the power cable and/or check the
rechargeable battery.
● Make sure that the switch on the switching power supply is switched
on.
● Check the connection of the interlock device.
Finger switch
If the "finger switch broken" error message is displayed:
● Check if the finger switch is chosen in the settings submenu.
● Check if the cable is properly connected to the control unit.
Foot control
If the foot control is not working:
● Check if the foot control is chosen in the settings submenu.
● Check the battery of the wireless foot control.
● Register the wireless foot control again.
Single-use fiber tip or therapy light guide
If the single-use fiber tip or therapy light guide missing error is
displayed:
● Make a visual check of the single-use fiber tip or therapy light guide
and its connector. If you see any damage (e.g. scratches) replace
the single-use fiber tip or therapy light guide with a new one.
● Check the connection of the single-use fiber tip or therapy light
guide.
● Check the proper assembly of the handpiece sleeve.
● Be sure that all operational steps have been carried out correctly.
Aiming beam
If there is no aiming beam:
● Check to see if the single-use fiber tip or therapy light guide or its
connector is damaged. If the single-use fiber tip or therapy light
guide is damaged, replace it with a new one.
● Check the connection of the single-use fiber tip or therapy light
guide.
● Check the proper assembly of the handpiece sleeve.
● Be sure that all operational steps have been carried out correctly.
If the aiming beam does not project a uniform circular pattern.
● Trim the end of the single-use fiber tip with the fiber cutter. Always
make the notch perpendicular to the optical fiber.