4 Installation
4.4 Initial start-up – procedure for proper assembly
Dentsply Sirona
Operating Instructions SiroLaser Blue
30
66 56 842 D3648
D3648.201.03.02.02 10.2018
2.
Open the sterile transport packaging tube of the EasyTip by tearing
off the seal label from the top of the packaging.
3.
Position the packaging tube with the EasyTip on the connection
socket by placing the funnel-shaped end of the packaging on the
handpiece.
4.
Press the packaging tube with the EasyTip against the
handpieceusing light pressure until the EasyTip perceptibly clicks
into place and is firmly seated.
5.
Remove the packaging tube from the handpiece and the connected
EasyTip.
6.
Please check whether the EasyTip is firmly seated on the
handpiece and perform a visual check to make sure that it has not
been damaged during shipment.
7.
Put the laser into operation by choosing any preset treatment. A
corresponding description is provided in chapter "Operation
[→ 36]".
WARNING
The aiming beam must not be aimed toward a person‘s eye. It
consists of an intensive light source even when set to a low power
level. Always wear protective goggles.
8.
As soon as the SiroLaser Blue is ready you can check to make sure
that the aiming beam illuminates evenly, i.e. projects a circular light
pattern (see adjacent picture).To do this, aim the EasyTip vertically
at a white background. If the beam shows no pattern or the beam
pattern is not illuminated evenly, the EasyTip may be damaged or
defective. In this case, return the EasyTip to your dental dealer so
that it can be replaced under warranty. Do not use any defective
EasyTips.