5 Operation
5.5 Main home screen
Dentsply Sirona
Operating Instructions SiroLaser Blue
54
66 56 842 D3648
D3648.201.03.02.02 10.2018
2.
Press the finger switch for at least 3 seconds while directing the
laser to the head of the power meter.
3.
Read the measured power from the display of your power meter.
4.
The unit will ask you whether the measured value is within the
tolerance of ±20%. Press 'Yes' on the screen if the measured value
is within the tolerance of ±20% Press `No` if it is outside the
tolerance.
5.
Repeat the procedure for all wavelengths.
If the calibration check is passed successfully, the message "Calibration
Test passed" will appear.
➢ Acknowledge with 'OK'.
If the laser shows an error message, please contact your local service.
5.5.3.2.9
Language and country settings
Language
Language is only available if "Country Settings" are set for NON-US.
English is preset and fixed for the US (the button is grayed-out).
➢ You have the choice of different languages. The currently set
language is greyed out. Select one and confirm by pressing 'OK'.
Ä
Language will be applied after confirmation.
NOTE
The language will be changed for all users.
Country settings
NOTE
The country setting for the US is pre-set.
WARNING
It's forbidden to change the country setting to Non US if you are
subject to US legal regulations. The use of the country setting Non
US is not authorized by the FDA.
For all users except US users:
➢ Change the pre-set country setting to Non-US and confirm the
selection.
Enter the pin code 3 3 3 4 and press 'OK '.
Ä
Now you will have access to the full range of pre-set
indications.
5.5.3.2.10
Service Menu
NOTE
Only authorized persons are allowed to enter the service menu. To
avoid misuse, it is necessary to enter the eight-digit pin code.