![Dentsply Sirona SiroLaser Blue Operating Instructions Manual Download Page 71](http://html1.mh-extra.com/html/dentsply-sirona/sirolaser-blue/sirolaser-blue_operating-instructions-manual_2483705071.webp)
Dentsply Sirona
Operating Instructions SiroLaser Blue
8 Maintenance and service
8.7 Replacing the batteries of the wireless foot control
66 56 842 D3648
D3648.201.03.02.02 10.2018
71
8.7
Replacing the batteries of the wireless foot
control
The wireless foot control is powered by two (2) AAA batteries
(commercially available).
When the battery is empty, select the finger switch in the settings
submenu "Activation device [→ 48]" for further operation of the
SiroLaser Blue.
The batteries can be changed by the user.
The housing of the wireless foot control must be opened to change the
battery. Touch a grounded metal part before opening the housing to
prevent damage to the PC board due to electrostatic discharge.
CAUTION
Prior to changing the batteries, switch the SiroLaser Blue off at the
main switch. This prevents accidental triggering.
Removing and replacing the batteries
+
–
+
–
1.
Remove the screws from the bottom of the foot control.
2.
Remove the cover and open the battery compartment.
3.
Pull the battery holder out of the battery compartment and replace
the batteries with new ones. Be careful to insert them with the
correct polarity (minus pole facing spring).
Assembling the foot control
1.
Place the battery holder back again in the battery compartment.
2.
Close the battery compartment with the cover.
3.
Screw tight the screws at the bottom of the foot control.
NOTE
After changing the batteries, start the SiroLaser Blue and check the
complete functionality of the foot control. In case the finger switch was
selected as preliminary activation device, it is necessary to re-select
the wireless foot control. It is not necessary to re-register the foot
control again at the SiroLaser Blue after changing batteries.