2 Safety information
2.8 Laser protective goggles
66 56 842 D3648
D3648.201.03.02.02 10.2018
11
2.8
Laser protective goggles
All persons present in the room e.g. patient, dentist, assistant must
always wear the appropriate laser protective goggles which are
delivered along with the SiroLaser Blue.
Check before use
Before using the laser protective goggles, please read and observe the
instructions for use provided by the manufacturer and attached to the
goggles in the case. Make sure that the laser protective goggles:
● are not damaged
● conform to standard EN 207 with the following protection levels:
– user goggles: LB5 (445 ± 5 nm / 970 -10/+15 nm)
and LB1 (660 ± 5 nm)
– patient goggles: LB5 (445 ± 5 nm / 970 -10/+15 nm)
and LB3 (660 ± 5 nm)
● are suitable for the correct wavelengths (labeled on the goggles)
These instructions apply particularly when using goggles supplied from
an outside source that are not included with the SiroLaser Blue.
Optical instruments
Never use optical instruments such as microscopes, eye loupes or
magnifiers together with the original protective goggles. Use of these
instruments with the protective goggles will no longer ensure sufficient
eye protection.
2.9
EasyTips and MultiTips
Optical fiber tips and connection socket
Make sure that no dust, dirt and foreign particles can enter the optical
fiber socket or the optical system. Never place your finger or any other
objects in the optical connectors. Otherwise the unit may be
permanently damaged.
When disconnecting the EasyTip or MultiTip from the SiroLaser Blue,
always cover the connection socket at the handpiece with the special
protection cap supplied. Make sure that the optical system is clean
before connecting the EasyTip or MultiTip.
The optical fiber must not be twisted inside the tube of the single-use
fiber tip (EasyTip). There is a risk of breakage of the tip.
Stop the laser activation of the SiroLaser Blue immediately if the
EasyTip or MultiTip is broken. Otherwise the tips may become hot.
EasyTips and MultTips must be checked for proper seating prior to each
use.
WARNING
Single-use fiber tips (EasyTip) must not be sterilized after usage.
They are disposable products and are not for re-use.