7 Cleaning, disinfection and sterilization
7.1 Cleaning
66 56 842 D3648
D3648.201.03.02.02 10.2018
63
7
Cleaning, disinfection and sterilization
Following treatment, switch off the SiroLaser Blue and disconnect the
power cable from the power supply.
NOTE
Wear gloves during these procedures.
Control unit, handpiece body, handpiece tube and foot control must be
cleaned and wipe-disinfected.
Dispose the single-use fiber tips.
The removable stainless steal handpiece sleeve, therapy light guides,
fibercutter and bending tool must be cleaned and sterilized.
CAUTION
Do not clean and disinfect the parts using a washer-disinfector! The
parts may be seriously damaged.
For the number of sterilization cycles, see chapter "Replacement of
parts subject to wear and tear [→ 72]".
7.1
Cleaning
NOTE
Manual cleaning must always be combined with disinfection.
Handpiece sleeve
1.
After disposing of the single-use fiber tip or removing the therapy
light guide, remove the handpiece sleeve from the handpiece body
by pressing the snap tab.
CAUTION
Danger of damage to the optical system
Reattach the protective cap to the optical system of the handpiece
immediately after removing the EasyTip or MultiTip. Do this before
conducting any cleaning process.
2.
Clean the handpiece sleeve with a suitable brush under running
water.
Therapy light guide (MultiTip)
➢ Clean the therapy light guide under running water (< 38 °C, water
must be at least drinking water quality).
NOTE
Never clean in an ultrasonic bath!