8 Maintenance and service
8.8 Replacement of parts subject to wear and tear
Dentsply Sirona
Operating Instructions SiroLaser Blue
72
66 56 842 D3648
D3648.201.03.02.02 10.2018
8.8
Replacement of parts subject to wear and tear
Check the following parts subject to wear and tear and replace where
applicable:
● Therapy light guides (change after 2,000 sterilization cycles or every
two years)
● Silicone key pad of the handpiece (change after 400 treatments/
sterilizations)
● Fiber cutter (change after 400 treatments/sterilizations or every two
years)
● Rechargeable battery (change after 1000 charging cycles or every
two years)
● Batteries in wireless foot control (change after 1 year)
For further information, see chapter "Cleaning, disinfection and
sterilization [→ 63]".
CAUTION
Only use parts from Dentsply Sirona, see "Spare parts [→ 26]".