4. Setting up the device
13
4.3.2. General device parameters
Arithmetic,
ArItH:
With this function the calculated value, not the measuring value, is shown in the
display. With
no
, no calulation is deposited. With
[P]
the selection is confirmed and the
device changes into menu level.
Rounding of display values,
round:
This function is for instable display values, where the display value is changed in
increments of 1-, 5-, 10- or 50. This does not affect the resolution of the optional
outputs. With
[P]
the selection is confirmed and the device changes into menu level.
Sliding average determination,
AVG:
Here, the number of the meterings that need to be averaged is preset. The time of
averaging results of the product of measuring time
SEC
and the averaged metering
AVG
.
With the selection of
AVG
in the menu level
DISPL,
the result will be shown in the display
and evaluated via the alarms.
Display time,
DISEC:
The display is set up with [
▲
] [
▼
]. Thereby it jumps until 1 second in increments of 0.1
seconds and until 10.0 seconds in increments of 1.0. With
[P]
the selection is
confirmed and the device changes into menu level.