4. Setting up the device
15
Special function [O]-key,
tASt.
4:
For the operation mode, special functions can be deposited on the
[O]-
Taste. This
function is activated by pressing the key. With
tArA
the device is set temporarily on a
parameterised value. The device acknowledges the correct taring with
oo0oo
in the
display.
Set.tA
adds a defined value on to the currently displayed value. Via
totAL
the
current value of the totaliser can be displayed for approx. 7 seconds, after this the
device switches back on the parameterised display value. If
tot.rE
is deposited, the
totaliser can be set back by pressing of the navigation keys [
▲
] [
▼
], the device
acknowledges this with
ooooo
in the display.
EHt.rE
deletes the min/max-memory. If
HOLD
has been selected, the moment can be hold constant by pressing the
[O]-
key,
and is updated by releasing the key.
Advice:
Hold
is activated only, if
HOLD
is selected
under parameter
DISPL
.
ActuA
shows the measuring value for approx. 7 seconds, after
this the device switches back on the parametrised display value. The same goes for
AVG
, here the sliding average values will be displayed. At
AL-1
…
AL-4
an output can be
set and therewith e.g. a setpoint adjustment can be done. If
no
is selected, the
[O]-
key
is without any function in the operation mode.
Assignment (deposit) of key functions,
tASt :
Via
totAL
the current value of the totaliser can be displayed for approx.7 seconds, after
this the device changes back on the parameterised display value. If
tot.rE
is
deposited, the totaliser can be set back by pressing of the navigation keys [
▲
] [
▼
],
the device acknowledges this with
ooooo
in the display. The configuration of
EHt.rE
deletes the MIN/MAX-memory. Under
ActuA
the measurand is shown for approx.
7 seconds, after this the display returns to the parameterised display value. If
AbS.UA
(absolute value) was selected, the display shows the value that has been measured
since voltage connection, without consideration of a previous taring. If
no
is selected,
the navigation keys are without any function in the operation mode.
Parameterisation level
Menu level
…