4. Setting up the device
10
4.3. Extended parameterisation
(Professional operation level)
Setting up the measuring range start/offset value,
offs
:
Enter the start/offset value from the smallest to the highest digit with [
▲
] [
▼
] and
confirm each digit with
[P]
. After the last digit the display switches back to the menu
level. If
Sens
was selected as input option, you can only select between
noca
and
cal
.
With
noca
, only the previously set display value is taken over, and with
cal
, the device
takes over both the display value and the analogue input value.
Setting the measuring range end value,
END
:
Set the end value from the smallest to the highest digit with [
▲
] [
▼
] and confirm each
digit with
[P]
. A minus sign can only be parameterised on the highest value digit. After
the last digit, the display switches back to the menu level. If
Sens
was selected as
input option, you can only select between
noca
and
cal
.
With
noca
, only the previously
set display value is taken over, and with
cal
, the device takes over both the display
value and the analogue input value.
Selection of the input signal,
tYPE:
Available as measuring input options are 0-50/-10 VAC or 0-5/-1 AAC signals as
works calibration (without application of the sensor signal) and
Se.50v, Se.10v, se.5a
and
se.1a
as sensor calibration (with the sensor applied). Select with [
▲
] [
▼
] and confirm
the selection with
[P]
.
Parameterisation level
Menu level
Menu group level
Menu level
4.3.1. Signal input parameters