8
4. Setting up the device
User code (4-digit number-combination, free available),
U.CodE
:
If this code was set (>0000), all parameters are locked for the user, if
LOC
has been
selected before under menu item
run
. By pressing
[P]
for 3 seconds in operation
mode, the display shows
COde
. The
U.Code
needs to be entered to get to the reduced
number of parameter sets. The code has to be entered befor each parameterisation,
until the
A.Code
(Master code) unlocks all parameters again.
The same applies to
LI-1
to
LI-2
!
Master code (4-digit number-combination, free available),
A.CodE
:
All parameters can be unlocked with this code, after
LOC
has been activated under
menu item
run
. By pressing
[P]
for 3 seconds in operation mode, the display shows
COde
and enables the user to reach all parametes by entering the
A.codE
. Under
run
the
parametrisation can be activated permanently by selecting
ULOC
or
ProF
, thus at an
anew pushing of
[P]
in operation mode, the code needs not to be entered again.
Function for threshold value undercut /exceedance,
Fu-1:
A limit value undercut is selected with
Louu
(for LOW = lower limit value), a limit value
exceedance with
High
(for HIGH = higher limit value). If e.g. limit value 1 is on a
threshold level of 100 and allocated with function
High
, an alarm is activated by
reaching of the threshold level. If the threshold value was allocated to
Low,
an alarm
will be activated by undercutting the threshold value, as long as the hysteresis is zero.
Hysteresis for limit values,
HY-1:
The delayed reaction of the alarm is the difference to the threshold value, which is
defined by the hysteresis.
Parameterisation level