If one or more of the control results are negative an error message will be displayed specifying the problem.
Here is an example of a negative test result where the camera has not been found – showing that communication with
the camera is probably faulty:
If “
Dilutor Error: X-Axis movement failed!
” appears – check that:
relevant connections are correct
belt tension is correct (ref. para.3.4.2.2..)
relevant motor is functioning correctly
relevant PCB is functioning correctly
If “
Dilutor Error: Y-Axis movement failed!
” appears – check that:
relevant connections are correct
belt tension is correct (ref. para.3.4.2.2.)
relevant motor is functioning correctly
relevant PCB is functioning correctly
If “
Dilutor Error: Z-Axis movement 1-2 probe failed!
” appears – check that:
relevant connections are correct
relevant motor is functioning correctly
relevant PCB is functioning correctly
If “
Dilutor Error: Blot Shaker failed!
” appears – check that:
relevant connections are correct
relevant motor is functioning correctly
relevant PCB is functioning correctly
If “
Dilutor Error: buffer system wash failed!
” appears – check that:
connections are correct
PCBs are functioning correctly
Peristaltic pumps are functioning correctly
If “
Dilutor Error : camera test failed!
”
or
“Dilutor Error: camera set-up test failed!” or
“Attention! The camera has not been found” a
re displayed check that:
connections are correct
PCBs are functioning correctly