3.3 DILUTOR UNIT
3.3.1. DESCRIPTION OF HOW IT WORKS
FUNCTION
The dilutor unit aspirates and dispenses samples, calibrators, controls
and all reagents, and also cleans
inside of Z
1
probe.
3.3.1.1.
Dilutor Unit
The DILUTOR UNIT is composed of following parts (clockwise direction):
1
Dilutor mechanism
Moves probe with step motor
2
2-way Electro-valve
For the aspiration of washing solution or for probe aspiration/dispensation
3
Syringe
To aspirate and dispense
Dilutor
movement
mechanism
INTERFA2
PCB
Syringe
SIDILOP
PCB
Valve support
Electro-valve