AP Blot– SERVICE MANUAL
SECTION 4
CALIBRATION PROCEDURES (XCALIB)
Doc. N° MDS-AP-26-00-02
Rev.01
Date: 02.04.2009
Pag. 73
4.3. DESCRIPTION OF MENU FUNCTIONS
Menu
Description
Main functions
Tank level sensors
Checks tank level sensors
Dispensation Test
Only enabled for APE set up
Dilution Test
Only enabled for APE set up
Movement Test
Starts test
Print Preview
Displays calibration settings
Prints out calibration settings
Turn on lights
Turns internal lights on/off
Close
Closes calibration
Exit
Exits programme
Dilutor
Calibrate probe position
Check and calibrate probe centering over work area.
Calibrate probe height
Check and calibrate maximum probe height
Calibrate LT
Check and calibrate liquid sensor levels (LT)
Calibrate shaking plane
Calibrate shaking plane
Calibrate syringe
Removing syringe, check and calibrate syringe, tip positions, check
aspiration and dispensation.
Calibrate SUL
Tests and calibrates SUL settings.
Test wash basin
Tests peristaltic pump and electro-valve workings.
Prime
Primes dilutor unit hydraulic circuit
Calibrate camera
Calibrate camera
Calibration settings
Displays calibration settings
Home
Returns all different instrument parts to Home position
Turn motors off
Turns off all motors
Reset
Resets all PCBs
Settings
Language
Setting language for display
Hardware configuration
Setting work area
Communication Port
To set up communication port
Serial Number
To set up /check serial number
?
Instrument information
Displays installed firmware version
Information additional components
Displays installed software component versions
About
Displays XCALIB version and manufacturer details