AP Blot– SERVICE MANUAL
SECTION 4
CALIBRATION PROCEDURES (XCALIB)
Doc. N° MDS-AP-26-00-02
Rev.01
Date: 02.04.2009
Pag. 79
4.7. MAIN FUNCTIONS
4.7.1. Tank level Sensors
From “
Main Functions
” pull down menu select “
Tank level sensors
” to check that
system is functioning correctly
The following window is displayed:
The condition of each tank will be automatically updated in some seconds and then displayed.
To check that sensors are correctly functioning:
1
Take off cap
2
Turn cap upside down to lower the level sensors (white cylinder)
3
Check condition of tanks "
Not full/not empty
" which is updated each time the sensor moves up or down.
If no sensor update occurs then check the following:
Check that level sensor floater
is not blocked or obstructed,
Check that level sensor cables are correctly connected to the PCBs
Replace Dilutor PCB (see ref. para. 3.8.3.).
4.7.2. Dispensation Test
Only enabled for APE set up
4.7.3. Dilution Test
Only enabled for APE set up