SECTION 3
MAINTENANCE PROCEDURES
Doc. N° MDS-AP-26-00-02
Rev.01
Date: 02.04.2009
Page 40
3.4.4.2.
Substituting Y-axis Movement Probe Motor
a) Remove top cover by undoing screws on top of device.
b) Remove connector which joins motor to AP22MOT PCB
c) Remove motor after loosening 4 fixing screws.
d) Place new motor in position and fix in place with 4 screws.
e) Check belt tension.
f)
Reconnect motor to AP22MOT PCB with connector
3.4.4.3.
Substituting Opto XGOLAV (X-axis movement) PCB
a) Remove top cover after undoing screws on top.
b) Remove connector attached to XGOLAV PCB.
c) To substitute it unscrew PCB from plate.
d) Place new PCB
in position and screw it to the plate with two fixing screws
e) Re-insert connector onto PCB.
f)
Check centering of X axis (see para. 4.8)
g) Replace top cover.
4 fixing
screws
2 fixing
screws
XGOLAV
PCB