AP Blot - SERVICE MANUAL
SECTION 3
MAINTENANCE PROCEDURES
Doc. N° MDS-AP-26-00-02
Rev.01
Date: 02.04.2009
Page 22
e) Remove 8 screws from bottle separation carter
f)
Remove 7 screws from aluminium shaking plane
g) Lift up and detach screws holding the shaking plane movement mechanism motor
h) NB: each time shaking plane support is moved, or removed calibrate shaking plane as described Para. 4.8.4.
Aluminium
shaking plane
Bottle separation
carter